Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-131
[:SOURce[n]]:VOLTage:LIMit[:AMPLitude]
This command selects the limit of voltage source.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
20 V
Usage
[:SOURce[n]]:VOLTage:LIMit[:AMPLitude] <value>
[:SOURce[n]]:VOLTage:LIMit[:AMPLitude]?
[:SOURce[n]]:VOLTage:LIMit[:AMPLitude]? DEFault
[:SOURce[n]]:VOLTage:LIMit[:AMPLitude]? MINimum
[:SOURce[n]]:VOLTage:LIMit[:AMPLitude]? MAXimum
[n]
Channel number: For single-channel instruments, the value is
1
.
<value>
0 to 20 V
Details
This command sets the voltage limit. The 2281S cannot source voltage levels that exceed this limit.
When you change the voltage limit to a value that is smaller than the set value of voltage, the voltage
setting will be forced to voltage limit.
Example
:VOLT:LIM 15
Set the voltage limit to 15 V
Also see
[:SOURce[n]]:<function>[:LEVel][:IMMediate][:AMPLitude]
(on page 7-131)