Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-47
:BATTery[n]:SIMulator:TVOLtage?
This command queries the real-time terminal voltage.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:BATTery[n]:SIMulator:TVOLtage?
[n]
Channel number: For single-channel instruments, the value is 1
Example
:BATT:SIM:TVOL?
The real-time terminal voltage is displayed.
Also see
None
:BATTery[n]:SIMulator:TVOLtage:PROTection[:LEVel]
This command sets the TVOLtage protection value in the battery simulator function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
21 V
Usage
:BATTery[n]:SIMulator:TVOLtage:PROTection[:LEVel] <value>
:BATTery[n]:SIMulator:TVOLtage:PROTection[:LEVel]?
:BATTery[n]:SIMulator:TVOLtage:PROTection[:LEVel]? DEFault
:BATTery[n]:SIMulator:TVOLtage:PROTection[:LEVel]? MINimum
:BATTery[n]:SIMulator:TVOLtage:PROTection[:LEVel]? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1.
<value>
0.5 to 21.0 V
Details
This command sets the TVOLtage protection value in the battery simulator function.
This value cannot be changed when the simulation is in progress. If you modify this value without
pausing the simulation, error 704, "Not permitted while battery model is running" occurs.