Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 2: General operation
077114601 / March 2019
2-15
Soft key area 2 for the battery test function
Soft key area 2 contains the settings shown in the following figure and explained in the table. To
change a setting, press the soft key to open a setting window. After configuration, close the window
by pressing the
ENTER
or
EXIT
key.
Figure 16: Soft key area 2 for battery test
Setting
Description
Measure AH
Specify whether to turn on the measuring A-H function or generate a model
after a battery test is complete.
Measure C
Specify whether to measure capacitance.
Script
Specify whether to run a script; you can select a specific index in the script to
run, how many measurement cycles to make, and you can view the results in
a table that can be saved to a USB flash drive.
Sample Int
Set the sampling intervals (in milliseconds) during the charging and
discharging.
Next
Switch to home screen 3.
Soft key area 3 for the battery test function
Soft key area 3 contains the settings shown in the following figure and explained in the table.
Figure 17: Soft key area 3 for battery test
Setting
Description
OVP
Set the voltage threshold value for overvoltage protection.
Graph
View the graph of the measurement. You can change the view of the graph
using the soft under the graph.
Data Sheet
View measurements in a spreadsheet format.
Buffer
Clear the data buffer.
Next
Switch to home screen 1.