Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-129
[:SOURce[n]]:<function>:PROTection[:LEVel]
This command sets the overvoltage protection setting of the source output.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
Voltage: 21 V
Current: 6.1 A
Usage
[:SOURce[n]]:VOLTage:PROTection[:LEVel] <value>
[:SOURce[n]]:VOLTage:PROTection[:LEVel]?
[:SOURce[n]]:VOLTage:PROTection[:LEVel]? DEFault
[:SOURce[n]]:VOLTage:PROTection[:LEVel]? MINimum
[:SOURce[n]]:VOLTage:PROTection[:LEVel]? MAXimum
[:SOURce[n]]:CURRent:PROTection[:LEVel] <value>
[:SOURce[n]]:CURRent:PROTection[:LEVel]?
[:SOURce[n]]:CURRent:PROTection[:LEVel]? DEFault
[:SOURce[n]]:CURRent:PROTection[:LEVel]? MINimum
[:SOURce[n]]:CURRent:PROTection[:LEVel]? MAXimum
[n]
Channel number: For single-channel instruments, the value is
1
.
<value>
The overvoltage or overcurrent protection level:
Voltage: 0.5 V to 21 V
Current: 0.1 A to 6.1 A
Details
Overvoltage protection restricts the maximum voltage level that the instrument can source.
Overcurrent protection restricts the maximum current level that the instrument can source. This
protection is in effect only for positive output voltages or currents.
When this attribute is used in a test sequence, it should be set before the turning the source on.
Example
:VOLT:PROT 20
:VOLT:PROT?
Set the voltage source protection to 30 V and query
the value.
Output: 20
Also see
None