Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-55
:BATTery[n]:TEST:SENSe:AH:ILIMit
This command sets and queries the maximum charging current during A-H measurement in the battery test
function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0.1 A
Usage
:BATTery[n]:TEST:SENSe:AH:ILIMit <value>
:BATTery[n]:TEST:SENSe:AH:ILIMit?
:BATTery[n]:TEST:SENSe:AH:ILIMit? DEFault
:BATTery[n]:TEST:SENSe:AH:ILIMit? MINimum
:BATTery[n]:TEST:SENSe:AH:ILIMit? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
0 to 6.1 A
Example
:BATT:TEST:SENS:AH:ILIM 2
Set the limit current value to 2 A.
Also see
None
:BATTery[n]:TEST:SENSe:AH[:LEVel]?
This command queries the real-time value of the battery capacity during A-H measurement.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:BATTery[n]:TEST:SENSe:AH[:LEVel]?
[n]
Channel number: For single-channel instruments, the value is 1.
Example
:BATT:TEST:SENS:AH?
The value of the battery capacity is returned.
Also see
None