Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-42
077114601 / March 2019
:BATTery[n]:SIMulator:LV
This command sets the low voltage (LV) in the battery simulator function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0 V
Usage
:BATTery[n]:SIMulator:LV <value>
:BATTery[n]:SIMulator:LV?
:BATTery[n]:SIMulator:LV? DEFault
:BATTery[n]:SIMulator:LV? MINimum
:BATTery[n]:SIMulator:LV? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
0.0 to 21 V
MINimum 0.000 V
MAXimum 21.000 V
DEFault 21.000 V
Details
When the output voltage reaches this value, the digital I/O pin, which is set as Digital Out and LV
action, outputs a level signal.
Example
:BATT:SIM:LV 0.1
Set the low voltage as 0.1 V
Also see
:BATTery:DIGital:LINE<line_index>:CONFigure
(on page 7-27)
(on page 7-28)