Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 4: Measurement optimization
077114601 / March 2019
4-3
Resolution controls the best possible combination of reading speed, accuracy, and front-panel display
digits.
You can independently adjust the front-panel display digits using the front panel or remote
commands. Setting the front-panel digits does not affect reading speed or accuracy. For details, refer
(on page 4-6).
If you change the resolution to 5.5 or 6.5, the instrument enables source delay and sets the delay
time to 1 ms.
Setting resolution
To set the resolution from the front panel:
1. Select the measurement function. The resolution setting will be applied to this function. For more
information, you can refer to
(on page 2-96).
2. Press the
HOME
key.
3. Press the
Next
soft key to switch to home screen 2.
4. Press the
Resolution
soft key.
5. Select one of the resolution values.
To set the resolution for the current, voltage, or concurrent function using SCPI commands:
SENse:CURRent:RESolution <n>
SENse:VOLTage:RESolution <n>
SENse:CONCurrent:RESolution <n>
Where
<n>
is:
4
= 4.5 digit resolution
5
= 5.5 digit resolution
6
= 6.5 digit resolution