Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Appendix C: Calibration and adjustment
077114601 / March 2019
C-11
Sink current calibration connections
For sink-current calibration, the following equipment is required:
Digital multimeter
Precision shunt resistors
Power supply
You use 2281S as an electronic load or a resistor to perform sink-current calibration.
You are required to change the shunt and load configuration for certain steps. Refer to the figures for
proper setup.
Figure 154: Sink-current calibration connection