Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-45
Example
:BATT:SIM:RES:OFFS 1
:BATT:SIM:RES:OFFS?
Set the offset value to 1 Ω and query the value.
Return: 1
Also see
None
:BATTery[n]:SIMulator:SAMPle:INTerval
This command sets and queries the sample interval in the battery simulation function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0.04s for 50 Hz
3.33333e-2s for 60 Hz
Usage
:BATTery[n]:SIMulator:SAMPle:INTerval <value>
:BATTery[n]:SIMulator:SAMPle:INTerval?
:BATTery[n]:SIMulator:SAMPle:INTerval? DEFault
:BATTery[n]:SIMulator:SAMPle:INTerval? MINimum
:BATTery[n]:SIMulator:SAMPle:INTerval? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
8e-5 to 0.48s for 50 Hz
6.66667e-5 to 0.5s for 60 Hz
Details
The sample interval is based on the power line cycles.
The shorter sample interval with the faster reading rate would increase reading noise and decrease
the number of usable digits, while the longer sample interval with the lower reading rate would provide
lower reading noise and more readable digits.
This value cannot be changed when the simulation is in progress. If you modify this value without
pausing the simulation, error 704, "Not permitted while battery model is running" occurs.
Example
:BATT:SIM:SAMP:INT 0.01
Set the sample interval to 0.01 s.
Also see
None