Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-40
077114601 / March 2019
Example
:BATT:SIM:CURR:LIM 1
Set the current limit to 1 A.
Also see
None
:BATTery[n]:SIMulator:CURRent:PROTection[:LEVel]
This command sets the current protection value in the battery simulator function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
6.1 A
Usage
:BATTery[n]:SIMulator:CURRent:PROTection[:LEVel] <value>
:BATTery[n]:SIMulator:CURRent:PROTection[:LEVel]?
:BATTery[n]:SIMulator:CURRent:PROTection[:LEVel]? DEFault
:BATTery[n]:SIMulator:CURRent:PROTection[:LEVel]? MINimum
:BATTery[n]:SIMulator:CURRent:PROTection[:LEVel]? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
0.1 to 6.1 A
MINimum 0.10 A
MAXimum 6.10 A
DEFault 6.10 A
Details
This command sets the current protection value in the battery simulator function.
This value cannot be changed when the simulation is in progress. If you modify this value without
pausing the simulation, error 704, "Not permitted while battery model is running" occurs.
Example
:BATT:SIM:CURR:PROT 2
Set the current protection to 2 A.
Also see
None