Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-139
[:SOURce[n]]:LIST:MCOMPlete
This command specifies the time when the instrument generates the measure complete signal with list enabled.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Saved settings
1
Usage
[:SOURce[n]]:LIST:MCOMPlete <type>
[:SOURce[n]]:LIST:MCOMPlete?
[n]
Channel number: For single-channel instruments, the value is 1
<type>
The time you can choose to generate :
STEP
: After each list step, the instrument will generate a measure complete
pulse; you can detect this pulse on digital I/O line 2; for more information,
refer to
(on page 3-22)
SWEEP
: After each list cycle, the instrument will generate a measure
complete pulse; you can detect the pulse on digital I/O line 2
Detail
When the list is running, the instrument allows you to check measure complete single to acquire the
running state. You can use this command to configure the measure complete signal. For example,
the following figure shows the different output of measure complete signal in a three-point list.
Figure 131: Step and sweep