Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-112
077114601 / March 2019
:OUTPut:PROTection:TRIPped?[<channel>]
This command queries whether overvoltage, overcurrent, overtemperature, or reverse fault protection has been
tripped.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:OUTPut:PROTection:TRIPped?
<channel>
The channel number (default value is CH1):
Channel 1: CH1
Channel 2: CH2
Channel 2: CH2
The
Details
There are six types of responses, including:
OCP indicates over-current protection occurred.
OVP indicates over-voltage protection occurred.
OTP indicates over-temperature protection occurred.
RVS indicates reverse sense fault detected.
OPS indicates open sense error protection occurred.
NONE indicates nothing occurred.
Example
:OUTP:PROT:TRIP? CH1
If the overvoltage protection circuit in channel 1 is
tripped, the query returns
OVP
.
If the protection circuit is not tripped, the query returns
NONE
.
Also see
None