Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-197
:TRACe[n]:TRIGger:VOLTage[:LEVel]
This command configures the voltage threshold for the level trigger.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0
Usage
:TRACe[n]:TRIGger:VOLTage[:LEVel] <voltageThreshold>
:TRACe[n]:TRIGger:VOLTage[:LEVel]?
<voltageThreshold>
The value of this parameter depends on the trace feed and relative status:
If
TRACe:FEED
is
CALC
, the voltage threshold ranges from -1.0E8 to
1.0E8
If
TRACe:FEED
is
SENSe
and Relative is off , the voltage threshold is
from 0 to 20
Example
TRAC:TRIG:VOLT 20
TRAC:TRIG:VOLT?
The value of the voltage threshold is set to 20.
Also see
:TRACe[n]:TRIGger:VOLTage:STATe
(on page 7-197)
:TRACe[n]:TRIGger:VOLTage:DIRection
(on page 7-196)
:TRACe[n]:TRIGger:VOLTage:STATe
This command determines whether the voltage level trigger function is enabled.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0 (OFF)
Usage
:TRACe[n]:TRIGger:VOLTage:STATe <state>
:TRACe[n]:TRIGger:VOLTage:STATe?
<
state
>
ON
or
1
: Enable the voltage level trigger function
OFF
or
0
: Disable the voltage level trigger function
Example
TRAC:TRIG:VOLT:STAT 1
TRAC:TRIG:VOLT:STAT?
The voltage level trigger function is enabled.
Also see
:TRACe[n]:TRIGger:VOLTage[:LEVel]
(on page 7-197)
:TRACe[n]:TRIGger:VOLTage:DIRection
(on page 7-196)