Appendix C: Calibration and adjustment
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
C-26
077114601 / March 2019
For 2281S, the expected values are shown in the following table.
Steps
Maximum voltage
Minimum voltage
9
6.06e-
2
V
–9.09e
-2
V
10
–0.2 V
–1.515 V
Step 9
1. Before starting this calibration step, configure the shunt and power supply as shown in the
following table.
Model
Shunt
Power supply
2281S
5.05 Ω
Constant current mode
5.0 V
0.001 A
2. Send the command to prepare for step 9:
:CALibration:PROTected:STEP3
3. Send the measured result from the digital multimeter (DMM) and shunt value using the following
command:
:CALibration:PROTected:STEP3:DATA <voltage>, <resistance>
4. Send the following command to query the operation status:
*OPC?
After all of the operations are completed, the instrument returns
1
.
Step 10
Before starting this calibration step, configure the shunt and power supply as shown in the following
table.
Model
Shunt
Power supply
2281S
5.05 Ω
Constant current mode
5.0 V
0.1 A
1. Send the command to prepare for step 10:
:CALibration:PROTected:STEP4
2. Send the measure result from the DMM and shunt value using the following command:
:CALibration:PROTected:STEP4:DATA <voltage>, <resistance>
3. Send the following command to query the operation status:
*OPC?
After all of the operations are completed, the instrument returns
1
.