Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-75
:BATTery[n]:OUTPut:PROTection:TRIPped?
This command queries whether over-protection has been tripped both in the battery simulator and battery test
function.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:BATTery[n]:OUTPut:PROTection:TRIPped?
[n]
Channel number: For single-channel instruments, the value is 1
Details
There are also six types of responses both in the battery test and battery simulator function, including:
OCP indicates overcurrent protection occurred.
OVP indicates overvoltage protection occurred.
OTP indicates overtemperature protection occurred.
RVS indicates reverse sense fault detected.
OPS indicates open sense error protection occurred.
NONE indicates nothing occurred.
Example
:BATT:OUTP:PROT:TRIP?
If the overvoltage protection circuit in channel 1 is
tripped, the query returns
OVP
.
If the protection circuit is not tripped, the query returns
NONE
.
Also see
None