Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Appendix C: Calibration and adjustment
077114601 / March 2019
C-31
Calibration constants
You can use the following command to read all the calibration constants in order of their index
number:
:CALibration:PROTected:DATA?
For 2281S, the default calibration constants are listed in the following table.
Index
number
Default
constant
value
Default
constant
tolerance
Calibration
steps
Error code
1
0.05 Ω
0.2 Ω
Step 5
Step 7
Step 11
Step 12
123, Step5 10A shunt value out of range;
128, Step7 10A shunt value out of range
168, Step11 Sink 1A shunt value out of range
172, Step12 Sink 1A shunt value out of range
2
5.05 Ω
10.1 Ω
Step 3
Step 4
Step 9
Step 10
116, Step3 100mA shunt value out of range;
120, Step4 100mA shunt value out of range
161, Step9 Sink 100mA shunt value out of range
165, Step10 Sink 100mA shunt value out of range
3
0.1
0.02
Step 1
110, Step1 Kmath error
4
0.3 V
0.25 V
Step 1
111, Step1 V-DAC zero DMM reading out of range
5
1.2946e
-2
2.589e
-2
Step 1
112, Step1 V-ADC zero collected data error
6
30 V
2 V
Step 2
113, Step2 V-DAC full scale DMM reading out of
range
7
1.2946
0.12946
Step 2
114, Step2 V-ADC full scale collected data error
8
0.003 V
0.0112 V
Step 5
122, Step5 I-ADC 10A zero DMM reading out of
range
9
8.571e
-3
1.714e
-2
Step 5
124, Step5 I-ADC 10A zero collected data error
10
8.5714e
-2
1.714e
-1
Step 5
125, Step5 I-ADC 1A zero collected data error
11
0.3 V
1.2 V
Step 7
127, Step7 I-ADC 10A full scale DMM reading out
of range
12
8.5714e
-1
8.571e
-2
Step 7
129, Step7 I-ADC 10A full scale collected data
error
13
8.5714e
-3
1.714e
-2
Step 6
126, Step6 I-DAC zero collected data error
14
8.2142e
-1
8.214e
-2
Step 8
130, Step8 I-DAC full scale collected data error
15
5.050e
-3
V 1.010e
-2
V Step 3
115, Step3 I-ADC 100mA zero DMM reading out of
range
16
1.428e
-2
2.857e
-2
Step 3
117, Step3 I-ADC 100mA zero collected data error
17
1.428e
-1
2.857e
-1
Step 3
118, Step3 I-ADC 10mA zero collected data error
18
0.505 V
1.01 V
Step 4
119, Step4 I-ADC 100mA full scale DMM reading
out of range
19
1.4285
1.428e
-1
Step 4
121, Step4 I-ADC 100mA full scale collected data
error
20
0
N/A
Step 9
160, Step9 Sink I-ADC 100mA zero DMM reading out
of range
21
0
N/A
Step 9
162, Step9 Sink I-ADC 100mA zero collected data
error
22
0
N/A
Step 9
163, Step9 Sink I-ADC 10mA zero collected data
error
23
0
N/A
Step 10
164, Step10 Sink I-ADC 100mA full scale DMM
reading out of range