Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-31
:BATTery[n]:MODel<modelindex>:<element>:APPend
This command adds values to the specified model.
Type
Affected by
Where saved
Default value
Command only
Instrument reset
Nonvolatile memory
Not applicable
Usage
:BATTery[n]:MODel<modelindex>:<element>:APPend <string>
[n]
Channel number: For single-channel instruments, the value is 1
<modelindex>
1 to 9
<element>
RESistance: resistance
Voc: open-circuit voltage
<string>
The setting values separated by commas; the maximum length of a string is
2048
Details
This command allows you to add values to the specified model. The new values are added to the end
of the existing values. You must set a total of 101 values in a model before executing it.
Example
:BATT:MOD1:RES:APPE "1.7,1.6,1.5"
Add the values of 1.7, 1.6, 1.5.
Also see
:BATTery[n]:MODel<modelindex>:<element>
(on page 7-30)