Section 3: Functions and features
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
3-44
077114601 / March 2019
When you run a list, you cannot change the following settings from front panel or using SCPI
commands. Sending the following commands causes an error 210, "Not permitted with list enabled."
Source settings that cause an error
Setting
Front-panel operations
SCPI commands
V-Set
MENU > Source
Settings > V-Set
HOME > V-Set
[:SOURce[n]]:<function>[:LEVel][:IMMediate][:AMPLitude]
I-Limit
MENU > Source
Settings > I-Limit
HOME > I-Limit
[:SOURce[n]]:<function>[:LEVel][:IMMediate][:AMPLitude]
Output delay
MENU > Source
Settings > Output
Delay
MENU > Source
Settings > Delay
Rise
MENU > Source
Settings > Delay Fall
(on page 7-109)
(on page 7-108)
Slew rise/fall
MENU > Source
Settings > Slew Rise
MENU > Source
Settings > Slew Fall
[:SOURce[n]]:VOLTage:SLEW:RISing
[:SOURce[n]]:VOLTage:SLEW:FALLing
OVP/OCP/
VMax
MENU > Source
Protect > Over
Voltage Protection
MENU > Source
Protect > Over
Current Protection
MENU > Source
Protect > Maximum
Voltage
HOME > Protect >
OVP
HOME > Protect >
OCP
HOME > Protect >
VMax
[:SOURce[n]]:<function>:PROTection[:LEVel]
[:SOURce[n]]:VOLTage:LIMit[:AMPLitude]
Source delay
MENU > Source
Delay > Source
Delay
MENU > Source
Delay > State
(on page 7-127)