Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-53
Details
This command specifies the range of open-circuit voltage for generating the battery model.
For generating a battery model, you need to specify the minimum Voc <low> and the maximum Voc
<high>.
If the <low> is greater than <high>, error 708, "Illegal model range setting" occurs.
If the range is not specified, the battery model will use the actual measured minimum and maximum
Voc.
Example
:BATT:TEST:SENS:AH:GMOD:RANG 2, 5
:BATT:TEST:SENS:AH:GMOD:RANG?
Set the Voc range as 2 to 5 V and query the range.
Return:
2, 5
Also see
None
:BATTery[n]:TEST:SENSe:AH:GMODel:SAVE:INTernal
This command saves the measurement results to the internal memory as a battery model.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:BATTery[n]:TEST:SENSe:AH:GMODel:SAVE:INTernal <modelindex>
[n]
Channel number: For single-channel instruments, the value is 1
<modelindex>
1 to 9
Details
This command saves the measurement results as a battery model to the internal memory. You can
save up to 9 battery models.
Example
:BATT:TEST:SENS:AH:GMOD:SAVE:INT 1
Save the measurement results as battery model 1.
Also see
:BATTery[n]:TEST:SENSe:AH:GMODel:RANGe