Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-22
077114601 / March 2019
:ARM[:SEQuence[n]]:SOURce
This command selects the control source for the arm layer.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
IMM
Usage
:ARM[:SEQuence[n]]:SOURce <source>
:ARM[:SEQuence[n]]:SOURce?
[n]
Channel number: For single-channel instruments, the value is
1
<source>
The trigger source for the arm layer:
Immediate triggering; no delay or gating in arm detection layer:
IMMediate
.
External triggering:
EXTernal
Manual triggering; the trigger signal is generated by pressing the
TRIGGER
key:
MANual
Bus triggering by the command
*TRG
:
BUS
Example
ARM:SOUR BUS
Select bus triggering as the trigger source
for the arm layer.
Also see
(on page 7-200)
(on page 3-30)