Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-161
Example
:STAT:QUES:INST:ISUM:ENAB #b0000000000001010
Sets the overvoltage protection (OVP)
and constant voltage (CV) bits of the
Questionable Instrument Summary
Enable Register using a binary value.
You could also send the decimal value
10
or the hexadecimal value
#hA
.
Also see
Questionable Instrument summary Event Register
(on page D-16)
:STATus:QUEStionable:INSTrument:ISUMmary:CONDition?
This command reads the Questionable Instrument Summary Condition Register of the status model.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:QUEStionable:INSTrument:ISUMmary:CONDition?
Details
This command reads the contents of the Questionable Instrument Summary Condition Register.
For details about interpreting the value of a register, see
Questionable Instrument Summary Event
(on page D-16).
Example
:STAT:QUES:INST:ISUM:COND?
Reads the Questionable Instrument
Summary Condition Register.
Also see
None