Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-44
077114601 / March 2019
:BATTery[n]:SIMulator:RESistance?
This command queries the real-time internal resistance of the battery simulator.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:BATTery[n]:SIMulator:RESistance?
[n]
Channel number: For single-channel instruments, the value is 1
Example
:BATT:SIM:RES?
The real-time internal resistance is displayed.
Also see
None
:BATTery[n]:SIMulator:RESistance:OFFSet
This command sets and queries the resistance offset value in the battery simulation function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0 Ω
Usage
:BATTery[n]:SIMulator:RESistance:OFFSet <value>
:BATTery[n]:SIMulator:RESistance:OFFSet
:BATTery[n]:SIMulator:RESistance:OFFSet? DEFault
:BATTery[n]:SIMulator:RESistance:OFFSet? MINimum
:BATTery[n]:SIMulator:RESistance:OFFSet? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
-
100 to 100 Ω
Details
This command sets the resistance offset value in the battery simulation function to cancel out the
resistance deviation caused by the wire connecting the battery and the device under test.
This value cannot be changed when the simulation is in progress. If you modify this value without
pausing the simulation, error 704, "Not permitted while battery model is running" occurs.