Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-196
077114601 / March 2019
:TRACe[n]:TRIGger:OFFSet
This command defines the number of points at which the instrument begins to log data in the buffer.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0
Usage
:TRACe[n]:TRIGger:OFFSet <value>
:TRACe[n]:TRIGger:OFFSet?
<value>
From
0
to
100
Example
TRAC:TRIG:OFFS 10
TRAC:TRIG:OFFS?
10 points before the level trigger activation are not cleared.
Also see
None
:TRACe[n]:TRIGger:VOLTage:DIRection
This command specifies whether the voltage rises above or falls below a specified level when the
level trigger is activated.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
RISE
Usage
:TRACe[n]:TRIGger:VOLTage:DIRection <direction>
:TRACe[n]:TRIGger:VOLTage:DIRection?
<
direction
>
The direction can be
RISE
or
FALL
Example
TRAC:TRIG:VOLT:DIR RISE
TRAC:TRIG:VOLT:DIR?
The level trigger will be activated when the voltage
rises above the preset threshold value.
Also see
:TRACe[n]:TRIGger:VOLTage:STATe
(on page 7-197)
:TRACe[n]:TRIGger:VOLTage[:LEVel]