Section 3: Functions and features
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
3-16
077114601 / March 2019
The current and voltage values in the buffer are different for the measurement function.
Measurement function
Current
Voltage
Concurrent (V + I)
Measured value for each
current reading
Measured value for each voltage reading
Voltage (V)
Current setting value
Measured value for each voltage reading
Current (I)
Measured value for each
current reading
Voltage setting value
For the battery test and battery simulation functions, the values of ESR and A-H are also saved in the
buffer.
To use the front panel to view the contents of the reading buffer:
1. Press the
MENU
key.
2. Under Views, select
Sheet
.
3. Select the button next to Data Sheet. The data is displayed.
4. If you want to view the statistics, select the button next to
Statistics
. The Statistics screen is
displayed.
5. Press the button next to
On.
The statistics results are displayed below the data table. To
configure the statistics function, see
Configuring the statistics applied to data in the buffer
page 3-18).
Figure 76: Data sheet in power supply function
6. Press the
HOME
key to return to the home screen.
Go to home screen 3. Press the
Data Sheet
soft key to view the buffer.