Section 3: Functions and features
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
3-18
077114601 / March 2019
Configuring the statistics applied to data in the buffer
The data buffer provides the following statistics:
Average
Minimum
Maximum
Peak-to-peak
Standard deviation.
When using the power supply function, you can request any of the statistics from the front panel or
using SCPI commands.
Specifying the statistics applied to the data in the buffer
Using the front panel:
1. Press the
MENU
key.
2. Under Measure, select
Data Buffers
.
3. Select the
Statistics
button. The Statistics window is displayed.
4. Enable the statistics you want to calculate.
Using the SCPI commands:
:CALCulate2:FORMat <name>
Where <
name
> can be any of the following options separated by commas:
MINimum
MAXimum
MEAN
SDEViation
PKPK
For example, you could enable minimum and maximum calculations by sending:
:CALCulate2:FORMat MIN, MAX