Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Appendix B: Verification
077114601 / March 2019
B-39
2. Set the voltmeter as follows:
a. Set to measure DC volts.
b. Set the current range to auto range.
c. Calculate the amperes (instead of volts) by multiplying the voltmeter reading result by 1/R.
3. Set the electronic load as follows:
a. Set to constant voltage.
b. Set to the specified voltage for your device under test (DUT) (see
(on page B-13)).
4. Set the device under test DUT to 0% of the full-scale output voltage.
5. Set the DUT to 100% of the full-scale output current.
6. Turn the DUT output on.
7. Enter the voltmeter reading in the
(on page B-13) table.
8. Change the AC power source output to the voltage specified in the following table.
DUT voltage selector switch
AC power source voltage
100
110 V
120
132 V
220
242 V
240
264 V
9. Enter the voltmeter reading in the Max line for your product in the
page B-13) table.
10. Change the AC power source output to the voltage specified in the following table.
DUT voltage selector switch
AC power source voltage
100
100 V
120
120 V
220
220 V
240
240 V
11. Enter the voltmeter reading in the Nom column for your product in the
(on page B-13) table.
12. Calculate the two values: Nom
– Min and Max – Nom. Enter the values in the appropriate
(on page B-13) table.
13. Select the largest of the two calculations from step 12 and enter the value in the Largest column
(on page B-13) table.
14. Turn the DUT output off.