Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 2: General operation
077114601 / March 2019
2-111
Math and statistics reset values
Setting*
Default value reset from
MENU > System >
Settings > Reset
Default value reset by the
(on page 7-7)
command or
page 7-175) command
MENU > Measure > Math > Type
:CALCulate[1]:<function>:FORMat
mx+b
MXB
MENU > Measure > Math > b (Offset)
:CALCulate[1]:<function>:KMATh:MBFactor
(on page 7-78)
+0.000e+0
0.0
MENU > Measure > Math > m (Gain)
:CALCulate[1]:<function>:KMATh:MMFactor
(on page 7-79)
+1.000e+0
+1.00
MENU > Measure > Math > Units
:CALCulate[1]:<function>:KMATh:MUNits
(on page 7-80)
X
"X"
MENU > Measure > Math > State
:CALCulate[1]:<function>:STATe
Off
0
(Off)
MENU > Measure > Data Buffers >
Statistics Format
All
"READ, UNIT"
MENU > Measure > Data Buffers >
Statistics Function
(on page 7-87)
Current
CURR
MENU > Measure > Data Buffers >
Statistics
Minimum On
Maximum On
Peak-Peak On
Mean On
Std Dev On
MIN, MAX, MEAN, SDEV,
PKPK
MENU > Measure > Data Buffers > Size
:TRACe[n]:POINts or :DATA[n]:POINts
100
100
MENU > Measure > Data Buffers >
Autoclear
:TRACe[n]:CLEar:AUTO or
:DATA[n]:CLEar:AUTO
(on page 7-186)
On
1
(On)
MENU > Measure > Data Buffers > Mode
:TRACe[n]:FEED:CONTrol or
:DATA[n]:FEED:CONTrol
(on page 7-191)
Always
ALW
(Always)
MENU > Views > Sheet > Statistics
Off
0
(Off)
* Some of the options listed in this table are only available for specific functions (power supply, battery test, or
battery simulation). If you do not see the option listed here on the menu, it may not be available for the function you
are using.