Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-56
077114601 / March 2019
:BATTery[n]:TEST:SENSe:AH:VFULl
This command sets and queries the full voltage during A-H measurement in the battery test function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0 V
Usage
:BATTery[n]:TEST:SENSe:AH:VFULl <value>
:BATTery[n]:TEST:SENSe:AH:VFULl?
:BATTery[n]:TEST:SENSe:AH:VFULl? DEFault
:BATTery[n]:TEST:SENSe:AH:VFULl? MINimum
:BATTery[n]:TEST:SENSe:AH:VFULl? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
0 to 20 V
Details
This command specifies the end voltage for charging. The voltage of the battery is equal or less than
this value.
Example
:BATT:TEST:SENS:AH:VFUL 2
Set the full voltage value to 2 V.
Also see
None
:BATTery[n]:TEST:SENSe:EVOC:DELAy
This command sets and queries the EVoc delay in the battery test function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0.1 s
Usage
:BATTery[n]:TEST:SENSe:EVOC:DELAy <value>
:BATTery[n]:TEST:SENSe:EVOC:DELAy?
:BATTery[n]:TEST:SENSe:EVOC:DELAy? DEFault
:BATTery[n]:TEST:SENSe:EVOC:DELAy? MINimum
:BATTery[n]:TEST:SENSe:EVOC:DELAy? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
0 to 0.1 s