Generating a battery model
To generate a battery model using the front panel:
1. Select the battery test function in the startup screen.
2. After charging is finished, go to the battery test home screen by pressing
HOME
on the front
panel.
3. Select
Measure AH
in the soft key area 3.
4. Select
Model: Generate
.
5. Set the Voc range for the battery model.
6. Name the battery model.
7. Select
Yes
to save the battery model. The 2281S can save up to nine battery models.
To generate a battery model using SCPI commands:
:STAT:OPER:INST:ISUM:COND?
:BATT:TEST:SENS:AH:GMOD:RANG 2.5, 3.7
:BATT:TEST:SENS:AH:GMOD:SAVE:INTE 1
These commands:
Query whether the measurement is done.
Set the range for generating the battery model.
Save the battery model to the internal memory.