Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 2: General operation
077114601 / March 2019
2-105
To set the maximum voltage limit using SCPI commands:
Send the
:SOURce:VOLTage:LIMit
command with the value of the limit. For example, to set the
voltage limit for the current source to 25 V, send the command:
:VOLTage:LIMit 25
See the command description for
[:SOURce[n]]:<function>:PROTection[:LEVel]
(on page 7-129) for
the full list of options.
Saving setups
You can save the present settings that you have defined for the 2281S to internal memory.
After the settings are saved, you can recall the settings. You can also set them to be the default
settings on power up.
Save a user setup
You can save a user setup to internal nonvolatile memory or an external USB flash drive.
To save a user setup to internal nonvolatile memory using the front panel:
1. Configure the 2281S to the settings that you want to save.
2. Press the
MENU
key.
3. Under System, select
Save/Recall
.
4. Select the button next to
Save
and press the
ENTER
key. The Save/Recall Location window is
displayed.
5. Select a location to save the setup and press
ENTER
. A keyboard is displayed.
6. Use the keyboard to enter the name of the setup.
7. Press the
ENTER
key. The setup is added to internal memory.
To save a user setup to external USB flash drive using the front panel:
1. Press the
MENU
key.
2. Under System, select
Save/Recall
.
3. Select the button next to
Save
and press the
ENTER
key. The Save/Recall Location window is
displayed.
4. Select
Save to USB
and press the
ENTER
key. A keyboard is displayed.
5. Use the keyboard to enter the name of the setup.
6. Press the
ENTER
key. The setup is saved to external USB flash drive.