Section 5: Application examples
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
5-14
077114601 / March 2019
8. Under
Measure
, select
A-H
.
9. Set the target voltage next to
V-Full
. If the battery works in the range of 3.8 V to 4.2 V, set this
value slightly higher than 4.2 V.
10. Set the maximum charging limit next to
Source I-Limit.
11. Set the end current to 10 mA next to
End condition
.
12. Return to the battery test home screen by pressing
HOME
on the front panel.
13. Select
Measure AH
in the soft key area 2.
14. Set the state to
On
. Then the charging starts with measurement result (voltage/current/Amp-Hour)
displayed on the screen.
15. Select
Model: Generate
after entering the soft key
Measure AH
when the charging is finished.
16. Set the Voc range for the battery model.
17. Enter a name to the battery model.
18. Select
Yes
to save the battery model. The 2281S can save up to nine battery models.
You can view the charging process by selecting
Graph
or
Data Sheet
in the soft key area 3 when
the charging is in progress.