Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-81
Also see
:CALCulate[1]:<function>:FORMat
(on page 7-77)
:CALCulate[1]:<function>:KMATh:MBFactor
(on page 7-78)
:CALCulate[1]:<function>:KMATh:MMFactor
(on page 7-79)
:CALCulate[1]:<function>:STATe
(on page 7-81)
:CALCulate[1]:<function>:STATe
This command enables or disables the math operations.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0 (OFF)
Usage
:CALCulate[1]:<function>:STATe <state>[,<channel>]
:CALCulate[1]:<function>:STATe? [<channel>]
<function>
The measurement function to which this setting applies:
Concurrent:
CONCurrent[:DC]
Current:
CURRent[:DC]
Voltage:
VOLTage[:DC]
<state>
Enable math operation:
ON
or
1
Disable math operation:
OFF
or
0
<channel>
The channel number (default value is CH1):
Channel 1: CH1
Channel 2: CH2
Details
When this command is set to on, the math operation specified by the math format command is
performed before completing a measurement.
Example
:CALC:VOLT:FORM MXB
:CALC:VOLT:KMAT:MMF 0.80
:CALC:VOLT:KMAT:MBF 50
:CALC:VOLT:KMAT:MUN 'C'
:CALC:VOLT:STATE ON
Set the math function for voltage measurements to mx + b.
Set the scale factor for voltage measurements to 0.80.
Set the offset factor to 50.
Set unit to 'C'
Enable the math function.
Also see
:CALCulate[1]:<function>:FORMat
(on page 7-77)
:CALCulate[1]:<function>:KMATh:MBFactor
(on page 7-78)
:CALCulate[1]:<function>:KMATh:MMFactor
(on page 7-79)
:CALCulate[1]:<function>:KMATh:MUNits
(on page 7-80)