Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-60
077114601 / March 2019
:BATTery[n]:TEST[:SOURce]:CURRent:PROTection[:LEVel]
This command sets and queries the protection threshold value of the output current in the battery test function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
6.1 A
Usage
:BATTery[n]:TEST[:SOURce]:CURRent:PROTection[:LEVel] <value>
:BATTery[n]:TEST[:SOURce]:CURRent:PROTection[:LEVel]?
:BATTery[n]:TEST[:SOURce]:CURRent:PROTection[:LEVel]? DEFault
:BATTery[n]:TEST[:SOURce]:CURRent:PROTection[:LEVel]? MINimum
:BATTery[n]:TEST[:SOURce]:CURRent:PROTection[:LEVel]? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
0.1 to 6.1 A
Details
This command specifies the protection threshold value of the current to be sourced to the battery
under test.
Example
:BATT:TEST:CURR:PROT 2
Set the protection threshold value to 2 A.
Also see