Section 2: General operation
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
2-104
077114601 / March 2019
Figure 68: Overprotection error dialog box
To clear an overprotection error using the front panel:
When the overprotection error dialog box is displayed, you can press the
ENTER
key to initiate the
trigger model.
To clear an overprotection error using a SCPI command:
When an overprotection error occurs, send the following command to clear the register and initiate
the trigger model:
:OUTPut:PROTection:CLEar
Maximum voltage limits
The maximum voltage limits prevent the instrument from sourcing a voltage over a maximum voltage
limit. This helps prevent damage to the device under test (DUT).
You cannot specify a voltage setting over the maximum voltage limit. If you set a maximum voltage
limit that is smaller than the voltage source value, the source value is automatically changed to the
maximum voltage limit.
Setting maximum voltage limits
To set the maximum voltage limit using the front panel:
1. Press the
MENU
key.
2. Under Source, select
Protect
.
3. Select
VMax
and set the limit.
4. Select
HOME
to return to the operating display.