Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-34
077114601 / March 2019
:BATTery[n]:MODel<modelindex>:ROW<rowindex>
This command sets the values of open-circuit voltage and resistance for a specified battery model.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Nonvolatile memory
Not applicable
Usage
:BATTery[n]:MODel<modelindex>:ROW<row_index> <string>
:BATTery[n]:MODel<modelindex>:ROW<rowindex>?
[n]
Channel number: For single-channel instruments, the value is 1
<modelindex>
1 to 9
<rowindex>
0 to 100
<string>
Values of Voc and resistance separated by comma
Details
This command provides a way to edit the model line by line.
Example
:BATT:MOD1:ROW0 "1.2, 1"
Set the Voc to 1.2.
Set the resistance to 1.
Also see