Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-140
077114601 / March 2019
Example
:LIST:MCOMP STEP
After each list step, the instrument generates a
measure complete pulse.
Also see
(on page 3-22)
[:SOURce[n]]:LIST:RCL
This command recalls a specific list.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Saved settings
0
Usage
[:SOURce[n]]:LIST:RCL <listNumber>
[:SOURce[n]]:LIST:RCL?
[n]
Channel number: For single-channel instruments, the value is
1
<listNumber>
The list number:
1
to
9
Details
Before you run a list, you need to recall this list. If you forget to recall a list, the last recalled list is
used. Sending the command
:LIST:RCL?
returns the active list number.
Example 1: Sweep list with BUS trigger (hold time type: TIME)
*RST
Define list
:LIST1:VOLT "1, 2, 3, 4, 5"
:LIST1:CURR "0.5, 1, 1.5, 2, 2.5"
:LIST1:DWEL "0.5, 0.5, 0.5, 0.5, 0.5"
:LIST:SAVE:INTe 1
:LIST:RCL 1
:LIST:MCOMP SWEEP
:LIST:HTIMe TIME
Define trigger model (sweep)
:ARM:SOUR BUS
:TRIG:COUN 5
Enable list mode
:LIST:STAT ON
:INIT
*TRG
Reset the instrument and stop the trigger model
Program voltage list 1 with 5 points
Program current list 1 with 5 points
Program dwell list 1 with 5 points
Save list 1 to nonvolatile memory
Recall list 1
Select sweep mode
Set hold time type to TIME
Select the BUS arm source
Set the trigger count to list points, no need to set the
sample count for TIME mode
Enable list mode
Start the trigger model
Run the lists
Hold time is the sum of the dwell time of 5 points
All the readings will be stored in data buffer