Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-194
077114601 / March 2019
:TRACe[n]:TRIGger:CURRent:DIRection
This command specifies whether the current rises above or falls below a specified level when the level trigger is
activated.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
RISE
Usage
:TRACe[n]:TRIGger:CURRent:DIRection <direction>
:TRACe[n]:TRIGger:CURRent:DIRection?
<
direction
>
The direction can be
RISE
or
FALL
Example
TRAC:TRIG:CURR:DIR RISE
TRAC:TRIG:CURR:DIR?
The level trigger will be activated when the current
rises above the preset threshold value.
Also see
:TRACe[n]:TRIGger:CURRent:STATe
(on page 7-195)
:TRACe[n]:TRIGger:CURRent[:LEVel]
(on page 7-194)
:TRACe[n]:TRIGger:CURRent[:LEVel]
This command configures the current threshold value for the level trigger.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0
Usage
:TRACe[n]:TRIGger:CURRent[:LEVel] <currentThreshold>
:TRACe[n]:TRIGger:CURRent[:LEVel]?
<currentThreshold>
The value of this parameter depends on the trace feed and relative
status:
If
TRACe:FEED
is
CALC
, the current threshold ranges from -1.0E7
to 1.0E7
If
TRACe:FEED
is
SENSe
and Relative is off , the current threshold
value is from 0 to 6.1
Example
TRAC:TRIG:CURR 1
TRAC:TRIG:CURR?
The current threshold value is set to 1.
Also see
:TRACe[n]:TRIGger:CURRent:STATe
(on page 7-195)
:TRACe[n]:TRIGger:CURRent:DIRection