Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-27
:BATTery:DIGital:LINE<line_index>:CONFigure
This command sets the digital I/O line function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
Line 1
– AUTO
Line 2
– MANual, MINPut
Line 3
– AUTO
Line 4
– MANual, MINPut
Line 5
– MANual, MINPut
Line 6
– MANual, MINPut
Usage
:BATTery:DIGital:LINE<line_index>:CONFigure <function>[,action]
:BATTery:DIGital:LINE<line_index>:CONFigure?
<line_index>
The digital I/O line: 1 to 6
<function>[,action
]
<function>
<action>
Description
AUTO
Automatically specify the pin state; this
function only applies to line 1 and 3:
Line 1: Trigger in; detects falling-edge,
rising edge, or either edge triggers as
input
Line 3: Protection output pulse; outputs
10μs TTL-compatible pulse when
OXP(OVP/OCP/OTP) or sense leads
reverse occurs
DIGIN
OUTPUTON
|
OUTPUTOFF
|
SSTARt
|
SSTOP
This function only applies to lines 2, 4, 5, 6;
SSTARt
|
SSTOP
are only available in TEST
mode; default <action> is
OUTPUTON
DIGOUT
FULL
|
EMPTy
|
CCCV
|
END
|
SSOurce
|
HV
|
LV
This function only applies to lines 2, 4, 5, 6;
FULL
|
EMPTy
are only available in SIM mode.
CCCV
|
END are
only available in TEST mode;
default <action> is
HV
MANual
MOUTH
|
MOUTL
|
MINPut
This function applies to all pins
Example
:BATT:DIGital:LINE1:CONFigure AUTO
:BATT:DIGital:LINE2:CONFigure DIGOUT,
CCCV
Set line 1 as AUTO.
Set line 2 as DIGOUT with CCCV.
Also see
None