Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-48
077114601 / March 2019
Example
:BATT:SIM:TVOLTAGE:PROT 12
:BATT:SIM:TVOLTAGE:PROT?
Set the TVOLtage protection value to 12 V and query
the value.
Return:
12
Also see
None
:BATTery[n]:SIMulator:VOC
This command sets the Voc value for the battery simulator function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Not saved
The maximum value in the
battery model
Usage
:BATTery[n]:SIMulator:VOC <value>
:BATTery[n]:SIMulator:VOC?
:BATTery[n]:SIMulator:VOC? DEFault
:BATTery[n]:SIMulator:VOC? MINimum
:BATTery[n]:SIMulator:VOC? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
The value range of the Voc: Minimum to Maximum
Details
This command sets the Voc value for the battery simulator function.
It allows the user to adjust the Voc value at any time even though the simulation is in progress.
After adjusting the Voc value, the SOC value will automatically update to the associated value
specified in the battery model.
Example
:BATT:SIM:VOC 4.5
Set the Voc value to 4.5 V.
Also see
(on page 7-46)