Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-115
:SENSe[n]:<function>:AVERage:COUNt
This command sets the number of measurements that are averaged when filtering is enabled for the specified
function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
10
Usage
:SENSe[n]:<function>:AVERage:COUNt <value>
:SENSe[n]:<function>:AVERage:COUNt?
:SENSe[n]:<function>:AVERage:COUNt? DEFault
:SENSe[n]:<function>:AVERage:COUNt? MINimum
:SENSe[n]:<function>:AVERage:COUNt? MAXimum
[n]
Channel number: For single-channel instruments, the value is
1
<function>
The measurement function to which this setting applies:
Concurrent:
CONCurrent[:DC]
Current:
CURRent[:DC]
Voltage:
VOLTage[:DC]
<value>
The number of readings required for each filtered measurement (
2
to
100
)
Details
The filter count is the number of readings that are acquired and stored in the filter stack for the
averaging calculation. When the filter count is larger, more filtering is done and the data is less noisy.
Example
SENS:CONC:AVER:WIND 0.1
Set the average filter window as 0.1%.
Also see
:SENSe[n]:<function>:AVERage[:STATe]
(on page 7-116)
:SENSe[n]:<function>:AVERage:TCONtrol
(on page 7-117)
:SENSe[n]:<function>:AVERage:WINDow
(on page 7-118)