Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-33
:BATTery[n]:MODel<modelindex>:<element>:STEPs?
This command queries the length of the specified model.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:BATTery[n]:MODel<modelindex>:<element>:STEPs?
[n]
Channel number: For single-channel instruments, the value is 1
<modelindex>
1 to 9
<element>
RESistance: resistance
Voc: open-circuit voltage
Details
This command queries the length of the specified model. The result of this command indicates the
number of values in the model.
Example
:BATT:MOD1:RES:STEP?
The number of values in the battery model is returned.
Also see
:BATTery[n]:MODel<modelindex>:<element>
(on page 7-30)
:BATTery[n]:MODel<modelindex>:<element>:APPend
(on page 7-31)