0x057A000A
4
Tube Too Tall for Cap
Piercing Sample Wheel.
Rack; Cup:
CAU:
Incompatible Tube Type
Message ID:
0x057A000A
Subsystem:
SRShuttle
This error will only occur when the CTS
thinks it has sensed a tube too soon. In
reality, it is impossible to put too tall of a
tube in because of the Gate Top and the
CTS cover. They sit too low to allow any
marginal tube height.
•
Make sure tubes are approved for use and in
appropriate rack for tube size. Refer to
.
•
Make sure rack is designated properly as a 100
or 75 mm rack in the Reserved Rack screen.
Refer to
Assigning or Reassigning Reserved
•
Make sure the tubes are seated properly in the
appropriate rack.
•
If tubes have been re-capped, make sure caps
are fully seated in the tube.
•
If error continues perform cap piercer tube
height alignment.
•
Contact Beckman Coulter Support Center for
additional information.
0x057A000E
4
Rack Height Not
Defined
CAU:
Rack Height Not Defined
Message ID:
0x057A000E
Subsystem:
SRShttl
CTS is installed and the sample rack
does not have a height defined.
•
Make sure the rack height is defined. Refer to
Figure 2.2, Reserved Racks/Obstruction
Detection Setup Dialog Box
.
0x057B0008
4
Offload Track Is Full
CAU:
Offload Track Is Full
Message ID:
0x057B0008
Subsystem:
SRLdr
This error will occur when the offload
trays sensor has been triggered. This
error is working as designed in most
cases.
•
Unload Racks from the Offload Tray.
•
If track is not full, make sure the switch at the far
right end of the track is not pressed in.
•
If the error continues, contact Beckman Coulter
Support Center.
0x05DC0070
4
Instrument Computer
Interprocess Error
CAU:
Boot Task
-
Message
Receive Error
Message ID:
0x05DC0070
Subsystem
: boot
Err no:
errno
Likely cause is a resource issue between
the Instrument computer processes.
There may be a mix-up during file
transfer or a file may be missing.
•
At the Red Message screen, select
•
If the problem continues after the instrument
has been shutdown, contact Beckman Coulter
Support Center for additional information.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......