A13914AF
8-21
User Defined Reagents
Expanded User Defined Chemistry Feature
8
7
On page 1 of the UDR setup, select the
Cal Save
check box.
ORDAC
ORDAC (Over Range Detection and Correction) allows a sample to be repeated with a smaller sample
volume if the first result exceeds the defined usable result range. Just like Beckman Coulter
chemistries, ORDAC can be turned on for automatic use through System Setup, or manually
requested in the Sample Program.
Enable/Disable UDR ORDAC
1
The instrument must be in
Standby
or
Stopped
mode.
2
Select
Setup
from the menu bar.
3
Select
15 User Defined Chemistries
.
4
Select the number to be defined.
OR
Type a number in the
Option Number
field.
5
Select
Define F1
.
6
Select
UDR+ F3
.
7
Go to page 2 of the UDR Setup, type the ORDAC sample volume, 2–40 μL. The volume entered
must be smaller than the initial sample volume.
8
Type the Lower and Upper Usable Result Range.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......