9-47
Maintenance
Three-Month Maintenance
9
5
Select
OK
when all components have been replaced in the syringe assemblies.
IMPORTANT
Always select
OK
to return the syringe plunger rods to the Home position (top) of the
syringe unit. Select
Cancel
to close the pop-up window without homing the syringes. If you selected
Cancel
from the pop-up window and the
Syringe Rod Replacement
screen is exited, a motion error
could occur.
6
Select
Close
to return to the Maintenance menu.
Prime the Syringes
1
Prime syringes 10 times. Do the following:
•
Select
Utils
from the menu bar,
•
Select
1 Prime
,
•
Select
CC F2
,
•
Select
Reagent
and/or
Sample Delivery Subsystem
,
•
Type
10
in the
Number of primes to repeat
field,
•
Select
MC F4
,
•
Select
Sample Delivery Subsystem
,
•
Select
Start Prime
.
2
Check the syringes and aspiration lines for air bubbles. If bubbles persist after priming:
Remove syringe and repeat "Install the Syringe," Steps 4–6 and "Prime the Syringes," Step 1.
Return to Home Position
1
Select
Instr Cmd
from the menu bar.
2
Select
1 Home
.
3
All instrument parts return to their home positions.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......