A13914AF
4-32
Reagent Load/Calibration
Reprint Calibration Reports
Reprint Previous Calibration Reports
1
Select
Rgts/Cal
from the menu bar.
2
Select
Options F6
.
3
Select
Print cal report
from the pull-down menu.
4
Select
Previous cal report(s)
from the
Print cal report
dialog box.
5
Select the desired Configured Chemistry name(s) from the
Previous cal report(s)
dialog box.
NOTE
When requesting reports of recent MC calibrations, select only one chemistry per report format.
This prevents duplicate copies. For example, if NA is selected, the report contains all five ISE
chemistries.
6
Select
Add
to move the chosen Chemistry(ies) to the Selected Chemistries area, then select
OK
.
7
Select 1, 2, 3, 4 or 5 most recent reports to print from the
Print Previous Cal Report
dialog box.
The report(s) automatically prints.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......