Maintenance
Six-Month Maintenance
12
When system has completed priming, select
Close
to exit the procedure.
CAUTION
Pinch hazard. Keep fingers clear of the ISE module as you lower it.
13
Lower the ISE module by releasing the locking pin. Be careful not to pinch tubing as the ISE
module is lowered.
14
Reinstall the ISE module cover.
15
Calibrate all the ISE chemistries. If the calcium electrode tip was replaced, proceed to Step 16.
IMPORTANT
Reference drift for the new electrodes may be observed during the first 4-6 hours after
electrode replacement. Results will
not
be compromised. Recalibrating the chemistry during this
time period helps stabilize the electrode.
16
Following calcium electrode tip replacement and calibration:
•
Program and run at least 20 replicates of serum samples for CALC.
•
Calibrate calcium.
Replace Glucose Sensor
The Synchron AccuSense glucose oxygen sensor requires no assembly. Each sensor has a gasket
positioned over the membrane tip. The sensor is discarded after six months usage on an instrument,
at which time a new sensor is installed.
Before you install the glucose sensor, record the sensor number found on the sensor. You must enter
this number into the Electronic Maintenance Log at the end of the procedure.
On DxC 800 Systems, the BUNm/UREAm module has an extra glucose sensor connector located next
to the BUNm/UREAm electrode connector. This extra glucose connector can be used to pre-
condition a new glucose sensor prior to installation.
Preparation Prior to Replacement (DxC 800 Systems)
For DxC 600 Systems, follow Steps 1–5 in the table below. Then proceed to Draining the Reaction Cup
for Sensor Maintenance or Replacement.
For DxC 800 Systems, plug the glucose sensor into the empty connector port, (labeled GLU on the
BUNm/UREAm module) 24 hours before glucose sensor installation minimizes sensor drift.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......