A13914AF
12-15
Troubleshooting Calibration and Result Errors
Error Codes and Definitions
12
"
Result Printed
"
CRITICAL HIGH
Exceeds Critical range high
"
Result Printed
"
CRITICAL LOW
Exceeds Critical range low
"
Result Printed
"
HIGH
Exceeds reference range high
"
Result Printed
"
LOW
Exceeds reference range low
"
Result Printed
"
TEMP ERR
TM
T
Temperature error
<X (X=lower inst range)
LT
Less than X - OIR LO
<X (X=lower reportable range)
LT
Less than X - ORR LO
>Y (Y=upper inst range)
GT
Greater than Y - OIR HI
>Y (Y=upper reportable range)
GT
Greater than Y - ORR HI
CAL FAILED
Calibration failed
CAL TIMED OUT
Calibration timed out
CHEM ABORTED
Too many retries
CHEM BYPASSED
Channel bypassed
CHEM NEEDS CAL
Chemistry needs calibration
CHEM NOT RUN
Chemistry not run
DAC ERROR
ISE DAC Failure
DIL1 LEV SENSE
Level sense error in diluent cartridge
DIVISION ERROR
ZD
Division Error Special calculation divisor is
zero
HI REAGENT
MC reagent level high
INCOMPLETE
AB
Calculation incomplete
INSUFF DIL1
Insufficient diluent reagent for scheduled
tests
INSUFF REAG
Insufficient reagent strength
INVALID CHEM
Invalid chemistry
LEV SENSE ERR
Low reagent fluid
LO REAGENT
MC reagent level low
MOTION ERR
Motion error
NO DIL1 ON SYS
No diluent reagent on the reagent carousel
NO REAG ON SYS
Reagent not on board
NO SAMP DETECT
Low sample fluid
NO SAMP ON SYS
Sample not on board
NOT ACCEPTED
Wrong state for sample program message
PENDING TEST
Not ready
PROBE OBSTRUCT
Probe obstruction detected
Table 12.9
Result Errors and Codes
(Continued)
Result Displayed on Report
Printed in
Remarks Area
Host
Code
Inst.
Code
Description
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......