A13914AF
3-33
System Setup Options
Auto Generation of Control
3
Auto Generation of Control
Introduction
When this feature is enabled, if a sample is loaded with a defined control ID, the instrument will
automatically
run any chemistry that is on-board and runnable for that control.
When Auto Generation of Control is enabled, the automatic Multiple Cartridge option is available.
With the Multiple Cartridge option enabled, the system automatically runs all cartridges on-board
and runnable for any chemistries for an Auto Generated control. Enabling Multiple Cartridge
without Auto Generation has no effect on the manual programming of control samples.
Setup
From the
Setup
screen, scroll down and select
29 Auto Generation of Control
to enable/disable either
feature.
The system must be in
Standby
,
Stopped, Startup, Instrument Down, or Homing
to make modifications.
The default is
disabled
.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......