12
0x139F0039
9
Reagent Cartridge has
less than 5 tests left:
Reagent:
reagent
Position:
position
Lot #:
lot num
Serial #:
serial num
STM:
Reagent container has
less than 5 tests left:
Reag:
chem
Lot No:
lot no
Serial No:
serial no
Pos:
position
Specified reagent cartridge has less than
5 tests remaining.
•
Prepare to load another cartridge if another is
not already on-board. Refer to
.
0x139F003A
9
MC Reagent container
has low volume:
Reag:
chem
Volume:
volume
Lot no:
lot no
Serial No:
serial no
STM:
Reagent container has
low volume remaining:
(
Reag:
chem
Volume:
volume
Lot no:
lot no
Serial No:
serial no
)
MC reagent container is getting low.
•
Begin preparations to load another container.
Refer to
Load/Unload Modular Chemistries (MC)
.
0x139F0068
7
CTS Tracking
Information can NOT be
shared between all
instruments. When you
move a tube to a
different instrument,
remove its cap. You will
be notified when full
CTS tracking is
restored.
STM:
TS-CTS Tracking System
disconnect detected
Message ID:
0x139F0068
This error will occur whenever 1 or more
instruments is no longer communicating
Cut Tube Tracking information with the
other networked instruments for any
reason.
IMPORTANT
TS-CTS stands for Thick-
Stopper, Closed Tube Sampling.
•
If a networked system has been shutdown and
rebooted, ignore this message. If not, check
power to all Microhubs and network cable
connections.
•
Select Status, then the CTS tab to determine
which instrument is no longer connected. Refer
to
.
CAUTION
To make sure a pre-cut tube is not cut a second
time, any pre-cut tube moved to the
disconnected system should have its cap
removed prior to loading.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......