A13914AF
12-17
Troubleshooting Calibration and Result Errors
Error Code – Definitions
12
Error Code – Definitions
Results Suppressed
RX ERR
RE
Reaction error
Results Suppressed
RX MAX DEV
RO
Reaction outlier (maximum deviation)
Results Suppressed
RX MEAN DEV
RN
Reaction noise (mean deviation)
Results Suppressed
RX NOISE
EA
A
Erratic ADC (Cup Chems only)
Results Suppressed
RXN RATE HI
RH
Reaction rate high
Results Suppressed
RXN RATE LO
RL
Reaction rate low
Results Suppressed
SAMP NOISE
DS
Sample signal noise (ISE only)
Results Suppressed
SAMP REF DRIFT
ES
Excessive reference drift – sample to
sample
Results Suppressed
SUB DEPL
SD
Substrate depleted
Results Suppressed
TEMP ERR
TM
T
Temperature error (MC only)
Results Suppressed
MOTOR ERR
HW
Stirrer Motor Error
TEST STOPPED
Test was aborted, system issue
TOO MANY TESTS
Too many tests on CC
Table 12.9
Result Errors and Codes
(Continued)
Result Displayed on Report
Printed in
Remarks Area
Host
Code
Inst.
Code
Description
Antigen Excess —
A chemistry failed antigen excess test for immunoprotein reagents. This flag may
indicate a contaminated cartridge.
Bad K/NA Value —
K or NA are required to calculate a CALC result. Should one or the other fail or have an
error, CALC cannot be calculated and this error is reported. Check your K and NA values for any error
conditions.
K is required to calculate a urine NA result. If K value has an error, urine NA cannot be calculated.
Blank Absorbance High/Low —
The mean absorbance measured during the reagent blank spin cycle
(read window) exceeds expected limits. The units are measured in absorbance and will characterize
reagent quality. This flag may indicate an issue with the reagent cartridge, reaction cuvette cleanliness or
sample integrity if chemistry is sample blanked.
Blank Noise (Mean Deviation) —
The average difference between the absorbance readings and the line
of regression is greater than allowed. This is a check for a constant rate during the blank spin cycle. The
flag may indicate possible issues with a photometer reading.
Blank Outlier (Maximum Deviation) —
A single blank absorbance data point obtained during the reagent
blank read window deviates more than allowed from the line of regression. The units are measured in
delta absorbance. Blank Maximum Deviation is a check for a constant rate during the blank spin cycles.
This flag may indicate possible issues with a photometer reading.
Blank Rate High/Low —
Rate measured during blank timing period is not within specifications. Units are
measured in delta absorbance/minute. This flag may indicate an issue with reagent contamination,
reaction cuvette cleanliness or sample integrity.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......